Spectrophotometer
Fluoroscence Spectrophotometer
ELICO's SL-174 spectrofluorometer is an instrument which takes advantage of fluorescent properties of some compounds in order to provide information regarding their concentration and chemical environment in a sample. A certain excitation wavelength is selected, and the emission is observed either at a single wavelength, or a scan is performed to record the intensity versus wavelength, also called an emission spectra
- Microprocessor based, User friendly, Menu Driven with Printer Interface
- Dual Monochromator Design.
- Four cell turret sample holder for Fluoroscence measurement.
- Precise settings of Wavelengths and Slit widths in both Excitation and Emission Monochromators.
- Compliance with IP/BP/USP standards.
- Supports Fluroscence Polarization modules (Single Cuvette sample holder)
- Absorbance, Fluorescence and Polarization.These three varieties of accessories are application oriented and are interchangeable.
- Optional PC compatibility with RS232C Interface and MS Windows® based Software for Data Acquisition, Processing, Storage, Retrieval & Interpretation of Data.
Applicable Industries
Academia & Research
Food & Beverages
TECHNICAL SPECIFICATIONS
SPECTRAL | |
Range (Ex & Em) | 200 to 900 nm |
Bandwidth | 2, 5, 10 & 20 nm (selectable) |
Readability | 0.1 nm |
Accuracy | ± 2 nm |
Repeatability | ± 0.2 nm |
Scanning Rate | 200, 400 & 600 nm/min (selectable) |
FLUOROMETRIC | |
Sensitivity | 0.1 ng/mL Quinine Sulphate is detectable with 20 nm bandpass in Ex & Em Monochromator. |
Raman Band of Water with 10nm bandpass at Ex WL 350 nm & Em WL 397 nm, S/N > 150. |
LIGHT SOURCE |
150 Watt Xenon Arc Lamp |
MONOCHROMATOR |
Czerny - Turner type with 1200 lines/mm Holographic Grating |
DETECTOR |
High Sensitive Photo Multiplier Tube (PMT) |
CONTROL | |
Standalone | Microprocessor |
Optional | Through Computer |
PC CONNECTIVITY |
RS 232C Interface |
SAMPLE ATTACHMENT | |
Standard | Motorised 4 position Cuvette holder for 10 mm Path Length Cuvettes. |
Optional | Polarizer module and Absorbance Module, Solid Sample Holder |
DATA PROCESSING | |
Microprocessor Mode (Standalone) | Spectrum (Excitation, Emission, Synchronous, PreScan), Concentration, Fluoroscence & GLP Compliance. |
PC Mode | Spectrum (Excitation, Emission , Synchronous, PreScan), Concentration, Fluoroscence, Time Scan, Quantum Yield Measurement & GLP Compliance. |
ACCESSORIES SUPPLIED | |
With Instrument | Fluoroscence free quartz cells of 10mm pathlength. |